三通道助磁(ci)直(zhi)阻(zu)測(ce)試儀(yi)(yi)不僅適用(yong)于(yu)配電(dian)變(bian)(bian)壓器(qi)(qi)(qi)、互感器(qi)(qi)(qi)、電(dian)抗器(qi)(qi)(qi)、電(dian)磁(ci)操(cao)作機構等感性試品,還適用(yong)于(yu)線材、開關(guan)觸點、繼電(dian)器(qi)(qi)(qi)觸點等阻(zu)性試品的測(ce)量。干式變(bian)(bian)壓器(qi)(qi)(qi)、非(fei)晶合金變(bian)(bian)壓器(qi)(qi)(qi),由于(yu)低(di)壓線圈采用(yong)銅箔繞制,電(dian)阻(zu)值低(di),對此本儀(yi)(yi)器(qi)(qi)(qi)將電(dian)流提升(sheng)至(zhi)10A,解決了低(di)值電(dian)阻(zu)測(ce)量的難題(ti)。
使用特定的(de)(de)電(dian)(dian)(dian)流(liu)源提供超過(guo)100A的(de)(de)測試電(dian)(dian)(dian)流(liu),并將其添加到(dao)要測試的(de)(de)環路電(dian)(dian)(dian)阻(zu)中。信(xin)號(hao)采樣電(dian)(dian)(dian)路用于收集環路電(dian)(dian)(dian)阻(zu)兩端的(de)(de)電(dian)(dian)(dian)壓(ya)和電(dian)(dian)(dian)流(liu)信(xin)號(hao)。獲(huo)得的(de)(de)信(xin)號(hao)通過(guo)放大(da)(da)器放大(da)(da),然(ran)后(hou)通過(guo)A/D轉換和微處理(li)器進行處理(li)/計(ji)(ji)算(suan),然(ran)后(hou)通過(guo)單片機進行計(ji)(ji)算(suan)/處理(li)以(yi)獲(huo)得所需的(de)(de)值,送到(dao)顯(xian)示(shi)設備(bei)中顯(xian)示(shi)出(chu)測量的(de)(de)電(dian)(dian)(dian)阻(zu)值。
在使用(yong)三通道助磁直阻測試儀(yi)的過程中如果出現(xian)故障,可通過以下方法解(jie)決(jue):
1、無法測出電阻(zu)值:造(zao)成該故障的原因多(duo)為以(yi)下兩點,分(fen)別是(shi):
1)開(kai)關電源壞了(le),無(wu)電流輸出;
2)設定值超出設備量程,程序(xu)無(wu)法識別指令(ling)。如果是電源(yuan)損(sun)壞,更換電源(yuan)即可。如果是超量程,重新啟(qi)動(dong)儀器,選(xuan)擇范(fan)圍內的電阻(zu)或電流進行(xing)測量即可。
2、開機就(jiu)燒保險:造成該故障也多為(wei)以下(xia)兩(liang)個(ge)原因:
1)5A-10A的(de)儀器(qi)出現這種情況,一(yi)般來說(shuo)是高壓將內置變壓器(qi)燒壞(huai),致使儀器(qi)無(wu)法正常使用(yong);
2)電流在20A以上(shang)的儀器出現這種情(qing)況,通常(chang)是開關電源(yuan)有問(wen)題。出現燒保(bao)險的情(qing)況,只(zhi)需(xu)及(ji)時更換(huan)變壓器或者(zhe)開關電源(yuan)即(ji)可正常(chang)使用。
3、使用(yong)過程中出(chu)現(xian)花屏現(xian)象(xiang):出(chu)現(xian)此故障的原(yuan)因(yin)有可(ke)能是(shi)內(nei)部排線(xian)接觸(chu)不良。關機檢查排線(xian)是(shi)否松動,加固排線(xian)即可(ke)解(jie)決問題。還有一種可(ke)能是(shi)由于電(dian)磁干擾導致屏幕(mu)顯示異常。如果(guo)是(shi)這個原(yuan)因(yin),只需讓廠家增加一個濾波(bo)電(dian)容即可(ke)。